Instruction 24-9460
46
Fyrite
®
 INSIGHT
Instruction 24-9460
47
Fyrite
®
 INSIGHT
6.3  O
2
 Sensor Replacement
Material Required:
 
• O
2
 Sensor (P/N 24-0788) 
Fyrite
®
 Rear Panel
Pry tab upward 
to remove the 
battery door 
from the unit.
CO 
Sensor
O
2
 
Sensor
Fig. 6.4
Procedure:
1.  After battery door is removed, remove 
connector tubing from both sensors. 
2.  Pull O
2
 sensor from its socket and 
remove the O
2
 cap.
3.  Properly dispose of the old sensor. En-
gage the nub on the new sensor within 
the slot on the cap's side and twist to 
secure the cap and sensor together.
4.  Install the cap and sensor unit by 
• Aligning the ribs on the sides of the 
sensor with the corresponding shape 
in the base. 
• Inserting the pins into the connec-
tors in the base. 
5.  Reattach tubing. 
6.2  Water Trap / Filter 
6.2.1 Emptying the Water Trap Chamber
The water trap chamber should be emptied after every test, or when the 
water condensate approaches the tip of the riser tube.
 
1.  Remove water trap chamber per Figure 6.3.
 
2.  Pour out liquid condensate, and then reassemble trap.
6.2.2 Replacing the Filter Element
Replace the filter element when it becomes visibly dirty or becomes satu-
rated with water.
Material Required:  
• Filter element, P/N 07-1644
 
• Small flat blade screwdriver
 
1.  Remove water trap chamber per Figure 6.3.
 
2.  Pry apart filter chamber using a small flat-blade screw 
  driver. Remove and discard old filter.
 
3.  Install new filter and reassemble filter chamber, making  
  sure that surfaces “A” and “B” contact each other.
 4. 
Reassemble 
trap.
Fig. 6.3
FILTER
ELEMENT
“A”
“B”
Pry apart 
filter chamber 
using a small 
flat-blade
screwdriver
inserted into 
notch
When reassem-
bling filter chamber, 
make sure surfaces 
“A” and “B” contact 
each other
WATER
TRAP
CHAMBER
FILTER
CHAMBER
Pull apart 
using a 
twisting
motion.
DO NOT 
rock back 
and forth!
O2 Cap
Sensor
Slot
Nub
Fig. 6.5