Bacharach H25-IR Manual De Usuario
H25-IR
Preparing for Operation
Instruction 3015-4342
17
2.5.6 Search Auto Zero
The SEARCH AUTO ZERO function allows the operator to select one of the following conditions when the
auto-zero function is disabled while in the Search mode:
NOTE:
This function works in conjunction with the setpoint value as entered in Section “2.5.5 Setpoint.”
ALWAYS ON
– The background gas level is always zeroed. If the probe is held over a gas leak, the
instrument will automatically zero-out the reading.
ALWAYS OFF
– The background gas level is never zeroed. If the probe is held over a gas leak, the
instrument will not zero-out the reading, similar in operation to the Measure mode.
BELOW SETPOINT
– The auto-zero function is ON when the leak-rate is below the setpoint value. If the
leak rate goes above the setpoint value, then the auto-zero function turns OFF, allowing the probe to stay
on the leak without the reading being zeroed-out.
on the leak without the reading being zeroed-out.
For example: If the Setpoint is 2 Oz/yr, then the auto-zero function is ON for leaks rates that are
<2 Oz/yr. If the leak rate is ≥2 Oz/yr, then the auto-zero function turns OFF.
<2 Oz/yr. If the leak rate is ≥2 Oz/yr, then the auto-zero function turns OFF.
BELOW 90% OF SP
– The auto-zero function is ON when the leak-rate is below 90% of the setpoint value.
If the leak rate goes above 90% of the setpoint value, then the auto-zero function turns OFF, allowing the
probe to stay on the leak without the reading being zeroed-out.
probe to stay on the leak without the reading being zeroed-out.
For example: If the Setpoint is 2 Oz/yr, then the auto-zero function is ON for leaks rates that are
<1.8 Oz/yr. If the leak rate is ≥1.8 Oz/yr, then the auto-zero function turns OFF.
<1.8 Oz/yr. If the leak rate is ≥1.8 Oz/yr, then the auto-zero function turns OFF.
BELOW 75% OF SP
– The auto-zero function is ON when the leak-rate is below 75% of the setpoint value.
If the leak rate goes above 75% of the setpoint value, then the auto-zero function turns OFF, allowing the
probe to stay on the leak without the reading being zeroed-out.
For example: If the Setpoint is 2 Oz/yr, then the auto-zero function is ON for leaks rates that are
<1.5 Oz/yr. If the leak rate is ≥1.5 Oz/yr, then the auto-zero function turns OFF.
<1.5 Oz/yr. If the leak rate is ≥1.5 Oz/yr, then the auto-zero function turns OFF.
BELOW 50% OF SP
– The auto-zero function is ON when the leak-rate is below 50% of the setpoint value.
If the leak rate goes above 50% of the setpoint value, then the auto-zero function turns OFF, allowing the
probe to stay on the leak without the reading being zeroed-out.
For example: If the Setpoint is 2 Oz/yr, then the auto-zero function is ON for leaks rates that are
<1 Oz/yr. If the leak rate is ≥1 Oz/yr, then the auto-zero function turns OFF.
<1 Oz/yr. If the leak rate is ≥1 Oz/yr, then the auto-zero function turns OFF.
2.6 Calibration
The H25-IR has been calibrated at the factory, and should not require frequent calibrations when
searching for gas leaks. When making measurements, however, the instrument should be calibrated on a
periodic basis to ensure that it meets its published accuracy specification.
periodic basis to ensure that it meets its published accuracy specification.
Instructions on how to calibrate the instrument are contained in Section 3 Calibration.
SEARCH AUTO ZERO
SETUP NO.1
<
ALWAYS ON
ALWAYS
OFF
BELOW
SETPOINT
BELOW 90% OF SP
BELOW 75% OF SP
BELOW 50% OF SP