Bacharach H25-IR Manual De Usuario

Descargar
Página de 60
H25-IR 
Preparing for Operation 
 
Instruction 3015-4342 
17
 
2.5.6 Search Auto Zero 
 
 
 
 
 
 
 
The SEARCH AUTO ZERO function allows the operator to select one of the following conditions when the 
auto-zero function is disabled while in the Search mode: 
NOTE:
 This function works in conjunction with the setpoint value as entered in Section “2.5.5 Setpoint. 
ALWAYS ON
 – The background gas level is always zeroed. If the probe is held over a gas leak, the 
instrument will automatically zero-out the reading.  
ALWAYS OFF
 – The background gas level is never zeroed. If the probe is held over a gas leak, the 
instrument will not zero-out the reading, similar in operation to the Measure mode. 
BELOW SETPOINT
 – The auto-zero function is ON when the leak-rate is below the setpoint value. If the 
leak rate goes above the setpoint value, then the auto-zero function turns OFF, allowing the probe to stay 
on the leak without the reading being zeroed-out. 
For example: If the Setpoint is 2 Oz/yr, then the auto-zero function is ON for leaks rates that are 
<2 Oz/yr. If the leak rate is ≥2 Oz/yr, then the auto-zero function turns OFF. 
 
BELOW 90% OF SP
 – The auto-zero function is ON when the leak-rate is below 90% of the setpoint value. 
If the leak rate goes above 90% of the setpoint value, then the auto-zero function turns OFF, allowing the 
probe to stay on the leak without the reading being zeroed-out.  
For example: If the Setpoint is 2 Oz/yr, then the auto-zero function is ON for leaks rates that are 
<1.8 Oz/yr. If the leak rate is ≥1.8 Oz/yr, then the auto-zero function turns OFF. 
 
BELOW 75% OF SP
 – The auto-zero function is ON when the leak-rate is below 75% of the setpoint value. 
If the leak rate goes above 75% of the setpoint value, then the auto-zero function turns OFF, allowing the 
probe to stay on the leak without the reading being zeroed-out.  
For example: If the Setpoint is 2 Oz/yr, then the auto-zero function is ON for leaks rates that are 
<1.5 Oz/yr. If the leak rate is ≥1.5 Oz/yr, then the auto-zero function turns OFF. 
 
BELOW 50% OF SP
 – The auto-zero function is ON when the leak-rate is below 50% of the setpoint value. 
If the leak rate goes above 50% of the setpoint value, then the auto-zero function turns OFF, allowing the 
probe to stay on the leak without the reading being zeroed-out.  
For example: If the Setpoint is 2 Oz/yr, then the auto-zero function is ON for leaks rates that are 
<1 Oz/yr. If the leak rate is ≥1 Oz/yr, then the auto-zero function turns OFF. 
 
 
2.6 Calibration 
The H25-IR has been calibrated at the factory, and should not require frequent calibrations when 
searching for gas leaks. When making measurements, however, the instrument should be calibrated on a 
periodic basis to ensure that it meets its published accuracy specification. 
 
Instructions on how to calibrate the instrument are contained in Section 3 Calibration
 
 
 
SEARCH AUTO ZERO 
SETUP NO.1 
        < 
ALWAYS ON 
 ALWAYS 
OFF 
 BELOW 
SETPOINT 
 
BELOW 90% OF SP 
 
BELOW 75% OF SP 
 
BELOW 50% OF SP