H25-IR 
Calibration 
Instruction 3015-4342 
21
 
3 Calibration 
When the instrument is supplied with an internal leak source, the operator can choose between two types 
of calibration: Internal or External. Internal calibration uses the instrument’s built-in leak source, while 
External calibration uses a customer supplied calibrated leak source.  
 
If the instrument is supplied without a leak source, then the option of selecting the calibration type is 
removed from the menu system. The only option for this model of instrument is to calibrate using an 
external leak source.  
 
 
3.1  Selecting Internal or External Calibration 
Use the Main Menu > CAL DATA > CAL TYPE function to select either Internal or External calibration. If 
External calibration is selected, then also use the EXT RATE and EXT GAS TYPE functions to enter the 
characteristics of the external leak source.  
 
Display the Calibration Data menu by first pressing both the ENT and ESC buttons on either the 
instrument’s front panel or Smart probe at the same time to display the Main menu. Next, use the Up and 
Down buttons to highlight “CAL DATA”, and then press ENT to display the Calibration Data menu. 
 
Main Menu:  
 
SYSTEM 
 LOG 
 SETUPS 
 CAL 
DATA 
 DIAGNOSTICS 
 
 
 
 
 
 
 
3.1.1 Internal Calibration Selection 
From the Calibration Data menu, use the Up and Down buttons to highlight “CAL TYPE”, and then 
press ENT to display its setup screen. Use the Up and Down buttons to highlight “INTERNAL”, 
and press ENT to make the selection and return to the Calibration Data menu. Press ESC as required to 
return to the Data Display screen. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Calibration Data Menu:  
 
CAL TYPE 
 EXT 
RATE 
 
EXT GAS TYPE 
 CAL 
FACTOR 
 RESTORE 
FACTORY 
CAL 
CAL TYPE 
INTERNAL 
 EXTERNAL