Oritron ICS-2100 Manual De Usuario

Descargar
Página de 288
ICS-2100 Ion Chromatography System
190
Doc. 065291-01 3/09
9. Feed the following liquid lines from the bottom of the holder through 
the tubing chase to the ICS-2100 front component panel.
TO SRS/AES REGEN OUT
TO INJ VALVE IN - P/ELUENT OUT
TO PUMP/DAMPER
10. Connect the lines to their respective locations:
Connect the line labeled 
TO SRS/AES REGEN OUT
 to the 
REGEN 
OUT
 port on the suppressor.
Connect the line labeled 
TO INJ VALVE IN - P/ELUENT OUT
 to 
port
P
 (
2
)
 on the injection valve.
Connect the line labeled 
TO PUMP/DAMPER
 to the pulse damper.
11. Go on to 
.
5.23.3 Removing the EPM and Reinstalling it in the New EGC 
Holder
1. Turn the EGC holder upside down to view the connections on the 
bottom of the holder.
2. Disconnect the following lines that connect from the EGC degas 
assembly inside the EGC holder to the EPM:
Disconnect the line labeled
 TO CR-TC ELUENT IN
 (red label) from 
the coupler that connects it to the line labeled
 TO CR-TC/EPM 
ELUENT IN 
(red label)
.
Disconnect the line labeled
 TO CR-TC ELUENT OUT 
(yellow label) 
from the coupler that connects it to the line labeled 
TO CR-
TC/EPM ELUENT OUT
 (yellow label).
Disconnect the line labeled 
TO CR-TC REGEN OUT
 (blue label) 
from the coupler that connects it to the line labeled 
TO CR-
TC/EPM REGEN OUT
 (blue label).
3. Remove the old EGC holder.