Smart Meter Verification flowchart: Running a test using the Chinese-language
display
Running a Smart Meter Verification test using the Chinese-language display
Figure 9-4:   
Stop?/Yes
. . . . . . . . . . . . . . . x%
Pass Verify
Abort Verify
Caution Verify
Fail
Abort
Sensor Rerun?
Yes
No
Correct condition
Run Verify
Continue Measure
Last Value
Select
Down
Down
Up
Select
Down
Results View/Yes?
Down
Down
Select
Select
Select
Select
Select
Sensor Abort?/Yes
Select
Up
Abort Type
Down
Down
Select
To Runcount
(see Results Read)
Test result
Pass
Exit
Down
Exit
Down
To Enter Meter Verify
Fault Value
Run a Smart Meter Verification test using ProLink II
1.
Choose Tools > Meter Verification > Run Meter Verification.
You may need to wait a few seconds while ProLink II synchronizes its database with
the transmitter data.
2.
Review the information presented on the screen, and click Next.
3.
Enter any desired information on the Test Definition screen, and click Next.
All information on this screen is optional.
4.
Choose the desired output behavior.
Measurement support
Configuration and Use Manual 
139