Abort A problem occurred with the meter verification test (e.g., process instability) or
you stopped the test manually. See 
each code, and possible actions you can take in response.
Smart Meter Verification abort codes
Table 9-3:   
Code
Description
Recommended actions
1
User-initiated abort
None required. Wait 15 seconds before
starting another test.
3
Frequency drift
Ensure that temperature, flow, and density
are stable, and rerun the test.
5
High drive gain
Ensure that flow is stable, minimize en-
trained gas, and rerun the test.
8
Unstable flow
Check factors that could cause process in-
stability, then rerun the test. To maximize
process stability:
Maintain a constant fluid pressure and
temperature.
Avoid changes to fluid composition,
e.g., two-phase flow or settling.
Maintain a constant flow rate.
13
No factory reference data for meter
verification test performed on air
Contact Micro Motion.
14
No factory reference data for meter
verification test performed on water
Contact Micro Motion.
15
No configuration data for meter ver-
ification
Contact Micro Motion.
Other
General abort
Repeat the test. If the test aborts again, con-
tact Micro Motion.
9.2.5
Schedule automatic execution of the Smart Meter
Verification test
You can set up and run a single test at a user-defined future time. You can also set up and
run tests on a regular schedule.
Manage scheduled test execution using the display
(standard option)
1.
Navigate to the Smart Meter Verification menu.
Measurement support
Configuration and Use Manual 
147