Emerson Model 1700 Manual De Usuario

Descargar
Página de 350
3.
To schedule recurring execution, specify a value for Hours Between Recurring Runs.
4.
To disable scheduled execution:
To disable execution of a single scheduled test, set Hours Until Next Run to 0.
To disable recurring execution, set Hours Between Recurring Runs to 0.
To disable all scheduled execution, click Disable Scheduled Execution.
Manage scheduled test execution using the
Field Communicator
1.
Navigate to the Smart Meter Verification menu:
Overview
 > Shortcuts > Meter Verification
Service Tools
 > Maintenance > Routine Maintenance > Meter Verification
2.
Choose Automatic Verification.
3.
To schedule a single test or the first test in recurring execution, specify a value for
Hrs Until Next Run
.
4.
To schedule recurring execution, specify a value for Set Recurring Hours.
5.
To disable scheduled execution:
To disable execution of a single scheduled test, set Hrs Until Next Run to 0.
To disable recurring execution, set Set Recurring Hours to 0.
To disable all scheduled execution, choose Turn Off Schedule.
9.3
Zero the flowmeter
Zeroing the flowmeter establishes a baseline for process measurement by analyzing the
sensor's output when there is no flow through the sensor tubes.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the flowmeter unless
one of the following is true:
The zero is required by site procedures.
The stored zero value fails the Zero Verification procedure.
Prerequisites
Before performing a field zero, execute the Zero Verification procedure to see whether or
not a field zero can improve measurement accuracy. See 
Important
Do not verify the zero or zero the flowmeter if a high-severity alarm is active. Correct the problem,
then verify the zero or zero the flowmeter. You may verify the zero or zero the flowmeter if a low-
severity alarm is active.
Measurement support
152
 Micro Motion
®
 Model 1700 Transmitters with Analog Outputs