Emerson Model 1700 Manual De Usuario
Need help?
If an error message appears:
•
Verify the Modbus address of the transmitter.
•
Ensure that you have specified the correct COM port.
•
Check the physical connection between the PC and the transmitter.
•
Increase or decrease resistance.
•
For long-distance communication, or if noise from an external source interferes with the
signal, install 120-
signal, install 120-
Ω ½-W terminating resistors in parallel with the output at both ends of
the communication segment.
•
Ensure that there is no concurrent Modbus communication to the transmitter.
D.3
Menu maps for ProLink III
Device Tools: Main
Figure D-9:
Using ProLink III with the transmitter
Configuration and Use Manual