Emerson Model 1700 Manual De Usuario

Descargar
Página de 350
4.3.1
Configure Volume Flow Type for gas applications
Display (standard)
Not available
Chinese-language
display
Not available
ProLink II
ProLink
 > Configuration > Flow > Vol Flow Type
ProLink III
Device Tools
 > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > Volume Flow Type > Standard Gas Volume
Overview
Volume Flow Type
 controls whether liquid or gas standard volume flow measurement is
used.
Procedure
Set Volume Flow Type to Gas Standard Volume.
4.3.2
Configure Standard Gas Density
Display (standard)
Not available
Chinese-language
display
Not available
ProLink II
ProLink
 > Configuration > Flow > Std Gas Density
ProLink III
Device Tools
 > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > Gas Ref Density
Overview
The Standard Gas Density value is used to convert the measured flow data to the standard
reference values.
Prerequisites
Ensure that Density Measurement Unit is set to the measurement unit you want to use for
Standard Gas Density
.
Procedure
Set Standard Gas Density to the standard reference density of the gas you are measuring.
Note
ProLink II and ProLink III provide a guided method that you can use to calculate the standard density
of your gas, if you do not know it.
Configure process measurement
Configuration and Use Manual 
33