Overview
You can set Update Period to control how frequently data is refreshed on the display.
Procedure
Set Update Period to the desired value.
The default value is 200 milliseconds. The range is 100 milliseconds to 10,000 milliseconds
(10 seconds).
5.1.5
Enable or disable automatic scrolling through the
display variables
Display (standard)
OFF-LINE MAINT
 > OFF-LINE CONFG > DISPLAY > AUTO SCRLL
Chinese-language
display
Offline Maintain
 > Configuration > Display > Auto Scroll
ProLink II
ProLink
 > Configuration > Display > Display Options > Display Auto Scroll
ProLink III
Device Tools
 > Configuration > Transmitter Display > General
Field Communicator Configure > Manual Setup > Display > Display Variable Menu Features > Auto Scroll
Overview
You can configure the display to automatically scroll through the configured display
variables or to show a single display variable until the operator activates Scroll. When you
set automatic scrolling, you can also configure the length of time each display variable is
displayed.
Procedure
1.
Enable or disable Auto Scroll as desired.
Option
Description
Enabled
The display automatically scrolls through each display variable as specified
by Scroll Rate. The operator can move to the next display variable at any
time using Scroll.
Disabled
 (de-
fault)
The display shows Display Variable 1 and does not scroll automatically. The
operator can move to the next display variable at any time using Scroll.
2.
If you enabled Auto Scroll, set Scroll Rate as desired.
The default value is 10  seconds.
Tip
Scroll Rate
 may not be available until you apply Auto Scroll.
Configure device options and preferences
62
 Micro Motion
®
 Model 1700 Transmitters with Analog Outputs