Emerson IFT9701 Manual De Usuario

Descargar
Página de 90
Model IFT9701 Transmitter Instruction Manual
67
IFT9701 Specifications
 continued
HA
R
T
 Me
n
u
 T
re
e
s
Retu
rn
 P
o
li
c
y
Op
tio
n
al Di
sp
la
y
Spec
ifications
20 to 30 VDC
6 watts typical, 14 watts maximum
Fused at 60 V/1.8 A Trip
Minimum startup voltage is 16 V at transmitter terminals
Maximum total resistance for wiring is 13
 ohms
At startup, transmitter power source must provide a minimum of 0.7 amp of short-term current
A.3
Environmental limits
A.3.1
Temperature
Operating
Optional local display may become difficult to read below +14 °F (–10 °C).
Storage
–40 to +185 °F (–40 to +85 °C) without optional LCD
–4 to +58 °F (–20 to +70 °C) with optional LCD
A.3.2
Process fluid temperature limits
Integrally mounted transmitter
–40 to +257 °F (–40 to +125 °C)
Remotely mounted transmitter
Refer to temperature limits for sensor.
Process fluid vs. ambient temperature
At their upper limits, process fluid temperature and ambient temperature restrict each other. The 
graphs provided below define the maximum recommended process fluid temperature based on 
maximum expected ambient temperature.