Honeywell SMV 3000 Manual De Usuario

Descargar
Página de 217
1/99
SMV 3000 Transmitter User’s Manual
103
9.4
Replacing Electronics Module or PROM
Module description
The electronics module used in the SMV 3000 transmitter is a two Printed
Wiring Assembly design that includes an integral mounting bracket, we
refer to the PWAs as Main PWA and Temperature or Daughter PWA as a
way to distinguish them.
PROM identification
The plug-in PROM on the main PWA is uniquely characterized to the
meter body of the given transmitter. For this reason, each PROM is given
a 10-digit identification number so you can verify that a replacement
PROM is the correct match for the given transmitter. The PROM
identification number is stamped on the nameplate on the transmitter’s
meter body and appears on a label under the PROM. You can also read the
PROM number using the SCT – See Section 8.2 in this manual for details.
Procedure
The procedure in Table 28 outlines the steps for replacing the electronics
module or the plug-in PROM. Since you must remove the electronics
module and PROM in either case, you can easily adapt the steps as
required.
Table 28
Replacing Electronics Module or PROM
Step
Action
1
Turn OFF transmitter power.
ATTENTION  We recommend that you remove the transmitter from
service and move it to a clean area before taking it apart.
2
Loosen end cap lock and unscrew end cap from electronics side of
housing.
Continued on next page