A&D MS-70 Manual De Usuario

Descargar
Página de 75
MS-70 / MX-50 / MF-50 / ML-50 
27 
8.1.2.
8.1.2.
8.1.2.
8.1.2.  Analyzing mode of the Automatic Mode
Analyzing mode of the Automatic Mode
Analyzing mode of the Automatic Mode
Analyzing mode of the Automatic Mode    
 
  When the change of moisture content per one minute is less than the preset value, 
the measurement is automatically completed. 
Range 
Preset Termination 
value to complete 
measurement 
MS-70 MX-50 MF-50 ML-50 
 
2.00 
%/min 
    
 
1.00 
%/min 
    
 0.50 
%/min 
 
 
Available  Available 
 0.20 
%/min 
 
Available 
range 
range 
 0.10 
%/min 
Available 
range 
(Factory setting)
 
(Factory setting)
 
 0.05 
%/min 
range 
(Factory setting) 
 
 
 0.02 
%/min 
(Factory setting) 
 
 
 
 0.01 
%/min 
 
 
 
 
 0.005 
%/min 
 
 
 
 
 0.002 
%/min 
 
 0.001 
%/min 
 
Unavailable item 
 
 
 
8.1.3.
8.1.3.
8.1.3.
8.1.3.  Analyzing mode of the Timer Mode
Analyzing mode of the Timer Mode
Analyzing mode of the Timer Mode
Analyzing mode of the Timer Mode    
 
  Sample is dried for a preset analysis time. 
 
  Drying Time   
1 minute to 480 minutes.   
 
  Setting interval : 
1 minute during 1 minute to 60 minutes. 
 
   
5 minutes during 60 minutes to 480 minutes. 
  
Factory 
setting: 
10 
minutes. 
 
 
 
8.1.4.
8.1.4.
8.1.4.
8.1.4.  Drying Program (
Drying Program (
Drying Program (
Drying Program (Heating Pattern and Drying Termperature)
Heating Pattern and Drying Termperature)
Heating Pattern and Drying Termperature)
Heating Pattern and Drying Termperature)    
Heating pattern
Heating pattern
Heating pattern
Heating pattern    
 
Standard Mode, 
Automatic Mode,   
Manual Mode 
Timer Mode 
Quick Mode 
Standard drying 
 
 
 
 
 
 
 
Ramp drying 
 
 
 
 
 
 
 
Step drying 
 
 
 
 
 
 
 
 
Quick drying 
 
 
 
 
 
 
"Temperature 1" can be set higher than "temperature 2" in step drying. 
ML-50, can select standard drying and quick drying only, does not display heating pattern.   
Temperature 
Temperature
Time 
Temperature
Temperature
Time1
Time2
50°C
50°C 
Time 
Stage 1 
Temperature1 
Time 
Temperature2
Stage 2
Temperature1
Time1
Temperature2
Stage 1
Stage 2
Time2
Temperature
200°C 
3 min.