Installation 
10/07 
Installation and Maintenance Manual 
 
 
3.5 Torque Recommendations 
For inserting a cell in metal fittings or bushings - 40 ft-lb maximum.  For inserting a cell in plastic fittings 
or bushings - 10 ft-lb maximum.  Always use pipe sealant (preferably Teflon tape). 
 
3.6 Dimension Drawings 
 
CE
LL
IN
1 1/2"
(38mm)
8 3/4"
(222mm)
14 1/2" max
(368mm)
3/4" Fitting
Allow 4 1/8" (105mm) for re
CE
LL
IN
1 1/2"
(38mm)
8 3/4"
(222mm)
14 1/2" max
(368mm)
3/4" Fitting
Allow 4 1/8" (105mm) for re moval of cell
moval of cell
3/4" NPT
(Male)
Flow In
Flow Chamber
1 1/2"
(38mm)
Octagon
3/4" NPT (Female)
Flow Out
1 1/2"
(38mm)
1 1/ 8"
(38mm)
Hexagon
3/4" NPT
(Male)
Flow In
Flow Chamber
1 1/2"
(38mm)
Octagon
3/4" NPT (Female)
Flow Out
1 1/2"
(38mm)
1 1/ 8"
(38mm)
Hexagon
2" min.
(51mm)
CEL
L
IN
a/n23342
Notes:
1. Mount cell and flow chamber horizontally as shown above with flow exit up to eliminate 
possible air gap around cell body.
2. If cell and flow chamber must be mounted vertically, attach a short length of tubing to flow 
exit as shown below and form a trap to ensure filling of flow chamber, especially at low flow.
 
Figure 3-1  Dimension Drawing for 055919 Flow Chamber