CF 26 
4/2001 
Transducer Excitation 
The bridge excitation supply voltage is set by SWB-1 and potentiometer E. Set SWB-1 ON for a fixed 10 Volts. This 
will provide the best temperature stability. The supply can be adjusted between 5 and 10 Volts by setting SWB-1 
OFF and adjusting potentiometer E. The supply will deliver up to 120mA current at any voltage setting to power up 
to four 350-ohm sensors. 
 
 
SWB-1
Excitation Voltage
ON
Fixed 10 Volts
OFF
Adjustable
 
 
 
 
 
SWA-1 Tare 
Range 
OFF 
-3 mV to +15 mV 
ON 
+15 mV to +25 mV 
 
SWA-2 Zero 
Current 
OFF 4 
mA 
ON 0 
mA 
 
SWA-4 SWA-5  Span 
Range 
ON 
ON 
40 mV to 50 mV 
OFF 
ON 
20 mV to 40 mV 
ON 
OFF 
10 mV to 20 mV 
OFF 
OFF 
5 mV to 10 mV