The *RST Command
Use this command to reset a device. It
Use this command to reset a device. It
initializes the device-specific functions in
the Counter.
The following happens when you use the
The following happens when you use the
*
RST command:
–
You set the Counter-specific functions to a
known default state. The *RST condition
for each command is given in the com-
mand reference chapters.
–
You disable macros.
–
You set the counter in an idle state (outputs
are disabled), so that it can start new oper-
ations.
The *CLS Command
Use this command to clear the status data
Use this command to clear the status data
structures. See ‘Status Reporting system’
in this chapter.
The following happens when you use the
The following happens when you use the
*
CLS command:
–
The instrument clears all event registers
summarized in the status byte register.
–
It empties all queues, which are summa-
rized in the status byte register, except the
output queue, which is summarized in the
MAV bit.