April 2004
4-50
Programming Site Data
Entering Correction Factors
The final option in the Program Menu, Correction Factors, allows 
you to enter and modify three correction factors: Reynolds 
Correction, Kinematic Viscosity and Calibration Factor. To enter 
the Correction Factors option, scroll to the Correction Factors 
entry on the Program Menu and press 
[ENTER]
. The screen 
appears similar to Figure 4-24 below.  To step through each 
parameter, press the [T] key. Refer to Figure A-7 on page A-7 of 
Appendix A, Menu Maps.
Figure 4-24:  Reynolds Correction Window
Entering Reynolds Correction
1. The first prompt asks if you want to enable the Reynolds 
Correction factor, a number based on the Kinematic Viscosity 
and flow rate of the gas. 
a. Use the  [
W
 ] and [
X
] keys to scroll to the appropriate 
radio button.
Reynolds Correction
Inputs
Correction Factors
Cancel
l
OK
Calibration Factor
Re Correction
On
Off
K/V x 10
-6
m
2
/s
Data Source
K/V Factors   
Single
Table
Edit Table
10