Displaying and Configuring Data
6-7
April  2004
The Measurement Option
On occasion, you might need to change the actual parameter 
measured in a given window.The Measurement option enables 
you to reconfigure the window with one of five categories of data 
source (velocity, volume, forward or reverse totalizer, or 
diagnostics) and appropriate English or metric measurement 
units.
To enter the Measurement option:
1. Press 
[SEL]
 from the Operate Mode window until you have 
reached the desired measurement. 
2. Press 
[ENTER]
 to open the Display Menu.
3. Be sure the Measurement option is highlighted, and press 
[ENTER]
The screen appears similar to Figure 6-5 on page 6-8. The left 
column displays the measurement data sources, while the right 
column displays English and metric measurement units (or, with 
the Diagnostics source shown, a list of diagnostics parameters).
4. Use the [
T
] and [
S
] arrow keys to reach the desired data 
source, and press 
[SEL]
5. Then use the [
T
] or [
S
] arrow keys to select the desired 
measurement unit (or diagnostic parameter).To speed 
scrolling, you can press the  [
X
] key to scroll down by a page, 
or the [
W
] key to scroll up by a page. Press 
[F3]
, OK, to 
confirm the entry.
6. The screen returns to Operate Mode, and displays any 
changes.