GE PT878GC Manual De Usuario

Descargar
Página de 308
April 2004
9-12
Servicing the PT878GC
Calibrating the Thickness Gauge Transducer
To zero the transducer offsets, go to the Zero tab. Press the [
W
 ] 
or [
X
] arrow key until you reach the Zero tab and press 
[ENTER]
The screen appears similar to Figure 9-9 below.
Figure 9-9:  The Zero Window 
1. Press the [
T
] arrow key to enter the window.
2. The first prompt asks you to choose between single and dual-
point calibration. Use the [
W
 ] and  [
X
] arrow keys to move 
to the appropriate radio button and press 
[ENTER]
Note: GE Panametrics recommends dual calibration. Dual-
point calibration is more reliable, but it requires two 
known thicknesses of the calibration material.Use single-
point calibration only if you know the sound speed in the 
calibration block material to a high degree of accuracy. 
See Appendix C for more information.
3. The next prompt asks you to enter the length of Block 1. (If 
you are using a GE Panametrics-supplied test block, the length 
is printed on the block.) Press 
[ENTER]
 to open the text box.
Display
Thickness Gauge Measure
Cancel
OK
Graph
Velocity
Zero
Material
Block 1 Length
Calculated
Current
Transducer Delay, 
0
         0
Single
Dual Calibration
0
mm Calibrate
µs