Servicing the PT878GC
9-15
April 2004
Calculating Velocity (Pipe Material Sound Speed) 
(cont.)
2. To enter the block length (here, the thickness of the sample as 
measured with a caliper or calibrated sample), press the [
T
arrow key and then press 
[ENTER]
 to open the Block Length 
text box. Use the numeric keys to enter the measured block 
length. Press 
[ENTER]
 to confirm the entry.
3. Prepare the sample for calibration by applying couplant B to 
its surface, and press 
[ENTER]
. The screen now appears 
similar to Figure 9-11 below.
4. Hold the transducer steady against the sample, and wait for the 
Set button to appear.
5. To calibrate the gauge, press the [
T
] arrow key to move to the 
Set button.
6. Press 
[ENTER]
 (Set button) to start the calibration sequence. 
Figure 9-11:  The Velocity Window, 
Displaying the Calculated and Current Values
Display
Thickness Gauge Measure
Cancel
OK
Velocity
Graph
Zero
Material
Block Length
Calculated
Current
1
Press Set to Commit Value
Set
22129.
74733.