April 2004
9-48
Servicing the PT878GC
Installing New Software (cont.)
1. As shown in Figure 9-30 on the previous page, the Flash 
Update option erases all instrument coding. 
Press 
[F2]
 (Cancel) to return to Operate Mode without 
erasing the program.
Press 
[F3]
 (OK) to confirm that you wish to erase the 
program.
2. The meter asks for confirmation. Repeat the options shown in 
Step 1 above. The screen on both the PC and the PT878GC 
now appears similar to Figure 9-31 below.
Figure 9-31:  Flash Reloading Window
3. The meter now asks if you wish to reload the flash memory 
via the infrared communications port. 
Press 
[F3]
 (Yes) (or C on the PC keyboard) to begin 
reloading via IrCOMM, or
Press 
[F2]
 (No) (or ESC on the keyboard) to stop the 
reloading procedure.
4. Be sure the infrared sensor on the meter has a clear path to the 
infrared adapter on the PC, and return to the PC 
Hyperterminal program.
GE Panametrics PCI Loader v3.0  2/26/04
[HW Rev3+]
Reload Flash via Infrared (Yes/No)?
Backup Battery: OK
Firmware CRC: Valid
IrOBEX          No           IrCOMM