GE PT878GC Manual De Usuario

Descargar
Página de 308
Diagnosis and Troubleshooting
10-3
April 2004
Table 10-1:  Error Messages
Error 
Message
Problem
Possible 
Cause
Action
E0
 No error.
Displays briefly 
after the dis-
play of another 
error message
None required. Mea-
surement is valid.
E1
Low Signal - 
Poor ultra-
sonic signal 
strength.
Broken cable. 
Flowcell prob-
lem. Transducer 
problem. Elec-
tronic failure.
Check transducer 
cable. See Flowcell 
Problems on page 10-
9 and Transducer 
Problems on page 10-
11. Check program-
ming and transducer 
spacing. Contact the 
factory if problem 
persists.
E2
Sound Speed 
Error (Sound 
speed is pro-
grammed 
using the Pro-
gram Menu 
Pipe option. 
as described 
in Chapter 4.)
Poor flow con-
ditions. 
Incorrect pro-
gramming. Bad 
transducer spac-
ing. Measure-
ment is very 
different from 
the pro-
grammed 
sound speed. 
Check sound speed 
against nominal sound 
speed. See Flowcell 
Problems on page 10-
9 and Transducer 
Problems on page 10-
11. Check program-
ming and transducer 
spacing.