Diagnosis and Troubleshooting
Displaying Diagnostic Parameters
As part of its measurement menu, the PT878GC offers a list of 
diagnostic parameters to aid in troubleshooting in the event of 
flowcell, transducer, or electrical problems. You can select any 
diagnostic parameter for display as a measurement as discussed 
in Chapter 6 on pages 6-7 and 6-8. Table 10-2 below shows all 
the available diagnostic parameters and ranges.
Table 10-2:   Diagnostic Parameters
Diagnostic
Parameter
 Displays
Good
Bad
∆T
Displays Delta T, or the differ-
ence between the upstream 
and downstream transit times.
N/A
Continuous 
large fluctu-
ations of 1 
micro sec-
ond or 
more.
Amplitude Up
Displays the value for the 
amplitude discriminator of the 
upstream transducer.
20-28
<20 or > 28.
Amplitude 
Down
Displays the value for the 
amplitude discriminator of the 
downstream transducer.
20-28 
fluctua-
tions
< 20 or > 28 
fluctuations
T Up
Displays the upstream transit 
time of the ultrasonic signal in 
micro seconds.
N/A
N/A
T Down
Displays the downstream tran-
sit time of the ultrasonic signal 
in micro seconds.
N/A
N/A
Gain/Up
Displays upstream gain in dB.
N/A
N/A
Gain/Dn
Displays downstream gain in 
dB.
N/A
N/A
CNTup
Displays the AGC DAC count 
for upstream gain setting.
N.A.
N.A.
CNTdn
Displays the AGC DAC count 
for downstream gain setting.
N.A.
N.A.
Soundspeed
(m/s or f/s)
Displays the measured sound 
speed of the fluid.
Check the sound speed 
of fluid in Sound Speeds 
and Pipe Size Data.