GE PT878GC Manual De Usuario

Descargar
Página de 308
April 2004
3-24
Installing DMP-1 Dampening Material with PI Fixture
1. Be sure the PI clamping fixture is installed onto the pipe with 
the transducers as described in the section Installing the PI 
Fixture and Transducers
 on page 3-12.
2. Approximate the axial distance from one transducer face to 
the other transducer face, as shown in Figure 3-27 below.
Figure 3-27:  Distance Between Transducer Faces
3. Lay the first strip of dampening material flat. Cut a strip to a 
width equal to the distance between transducer faces (the 
distance illustrated in Figure 3-27 above).
4. Wrap this strip around the pipe in the space between the 
transducers.
5. Lay the second strip of dampening material flat. Then cut the 
strip into two pieces, each 4.5-in. (12 cm) wide.
6. Wrap each of these strips around the pipe on the outside edge 
of the clamping fixture, one upstream and one downstream. 
The completed installation should appear similar to Figure 
3-28 on the next page.