April 2004
4-16
Programming Site Data
Path Parameters for Clamp-On Transducers
The PT878GC path menu includes two parameters for clamp-on 
transducers:
Traverses
Spacing
1. The PT878GC first prompts for traverses, the number of times 
the ultrasonic signal crosses the pipe. 
a. Press 
[ENTER]
 to open the drop-down list of traverse 
numbers.
b. Scroll to the appropriate number. 
c. Press 
[ENTER]
 to confirm the entry.
2. The next prompt displays the spacing of the transducers, as 
calculated from the information entered.
a. Record this number and use it to space the transducers. 
(Use Chapter 3 to assist in installing transducers on the 
pipe.) Press the [T] key to scroll to the next parameter.
b. If necessary, you can overwrite the spacing to match the 
actual physical spacing of the transducers. (Do not change 
the spacing by more than ±10% from that calculated by the 
meter.) Press 
[ENTER]
. Use the numeric keys to enter the 
desired value, and press 
[ENTER]
 to confirm the entry. If 
you enter an invalid entry, the PT878GC rejects the entry 
and displays an error message.
Note: GE Panametrics does not recommend using a spacing 
other than the one calculated by the PT878GC.
After you enter the spacing, pressing the [
S
] key returns the 
prompt to the Traverses box, and then to the Path tab at the top of 
the screen. 
To confirm the entries and return to Operate mode, press 
[F3]
 
(OK). 
To leave the window without confirming the entries and return 
to Operate mode, press 
[F2]
 (Cancel) or the 
[ESC]
 key.