April 2004
4-28
Programming Site Data
Entering Skan/Measure Integrate Parameters
To enter parameters for the Skan/Measure Integrate mode in the 
Skan/Measure option, press the [X] arrow key once to reach the 
Integrate tab. Then press 
[ENTER]
 to open the Integrate window, 
which appears similar to Figure 4-12 below.
Figure 4-12:  The Integrate Window in the skan/
Measure Option
1. The first prompt, A Divisor, is used in the calculation of the 
Measure Mode integrated threshold level and is not normally 
changed. The default value for this parameter is 2.5, and 
values from 0.1 to 10 are acceptable.
a. Press 
[ENTER]
 to enter the text box.
b. Use the numeric keys to enter the recommended value.
c. Press 
[ENTER]
 to confirm the entry.
2. Next, enter the Noise Level  ( -Offset % of Peak) from 0 to 
95%, and press 
[ENT]
. (The default value is 20%.)
a. Press 
[ENTER]
 to enter the text box.
b. Use the numeric keys to enter the recommended value.
c. Press 
[ENTER]
 to confirm the entry.
Integrate
Skan/Measure
Cancel
l
OK
All
A Divisor
Noise Level
Count
Correlate
2.5
Measure Only
     5
%