GE PT878GC Manual De Usuario

Descargar
Página de 308
April 2004
4-32
Programming Site Data
Entering Skan/Correlation Parameters
To open the Skan/Correlation window in the Skan/Measure 
option, press the [X] arrow key three times (from the All tab) to 
reach the Correlate tab, and press 
[ENTER]
. The screen appears 
similar to Figure 4-14 below.
 
Figure 4-14:  The Correlate Window in the Skan/
Measure Option
To identify the peak of the received signal, the PT878GC uses the 
“Threshold” method, in which the peak is identified as the point 
where the signal crosses a threshold that is a percentage of the 
maximum signal detected. 
1. The first prompt, Min Threshold, asks for the minimum 
percentage of peak used to measure transit time. 
a. Press 
[ENTER]
 to enter the text box.
b. Use the numeric keys to enter the recommended value.
c. Press 
[ENTER]
 to confirm the entry.
Integrate
Skan/Measure
Cancel
l
OK
All
Min Threshold
Percent Peak
%
Count
Correlate
-100
Measure Only
100
-40
%
%
Max Threshold