presence by entering their levels on the Test Re-
cord.
9. Set F1 to 1.6 GHz. Measure the worst case non-
harmonic signal for the 1.6 GHz carrier and rec-
ord its presence by entering its level on the Test
Record.
10. Change the Spectrum Analyzer setup as follows:
a. Span: 10 MHz/div
b. CF: 1.6 GHz
c. RBW: 1 MHz
11. Adjust the Spectrum Analyzer Reference Level
control to place the signal at the top of the screen
graticule.
12. Change the Spectrum Analyzer CF first to
3.2 GHz, then to 4.8 GHz. Compare the harmonic
levels with the signal level at 1.6 GHz. Measure
the harmonic levels and record them on the Test
Record.
3-16
682XXB/683XXB MM
PERFORMANCE
SPURIOUS SIGNALS TEST: RF OUTPUT SIGNALS
VERIFICATION
£2 GHz (£2.2 GHz for 68XX5B MODELS)