Chapter 5, Advanced Operations
Passing Control
Requesting Control using IBASIC 
The Test Set has the capability to request control of the bus from the Active 
Controller from a running IBASIC program using the IBASIC command 
EXECUTE ("REQUEST_CONTROL"). When the EXECUTE 
("REQUEST_CONTROL") command is executed from a running IBASIC 
program, the Request Control bit, bit 1, of the Test Set’s Standard Event Status 
Register is set to the TRUE, logic 1, condition. The Active Controller detects the 
request in the Test Set’s Standard Event Status Register either as a result of an 
SRQ indication by the Test Set or by some polling routine which periodically 
checks bit 1 of the Standard Event Status Register of all potential controllers on 
the bus. The Active Controller would then send the Test Set the address to which 
the Test Set is to later pass control using the *PCB Common Command. The 
Active Controller would then pass control to the Test Set. 
Pass Control Examples 
The following examples illustrate how pass control could be implemented in two 
of the common Test Set operating configurations: 
1. Test Set controlled by an external controller, and 
2. Test Set running an IBASIC program with an external Controller connected to GPIB 
bus. 
Passing Control While the Test Set is Controlled by an External Controller 
This example illustrates passing control between the Test Set and an external 
controller while the Test Set is being controlled by the external controller. In this 
mode the Test Set is NOT configured as the System Controller. Generally 
speaking, in this mode of operation the Test Set is considered just another device 
on the GPIB bus and its Controller capabilities are not used. However, it may be 
desirable, under certain conditions, to print a Test Set screen to the GPIB printer 
for documentation or program debugging purposes. With manual intervention it is 
possible to have the Active Controller pass control to the Test Set, have the 
operator select and print the desired screen, and then pass control back to the 
formerly Active Controller. The following steps outline a procedure for 
accomplishing this task. The example is based upon having an HP
®
 9000 Series 
300 Workstation as the external controller connected to the Test Set through the 
GPIB bus. Further, it assumes that the GPIB interface in the HP
®
 9000 Controller 
is set to the default select code of 7 and address of 21.