Tektronix TDS460A Manual De Usuario
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–323
TRIGger:MAIn:VIDeo? (Query Only)
Option 05 Only
Returns the main video trigger parameters.
Trigger
TRIGger:MAIn:VIDeo?
TRIGger
:
:
VIDeo
MAIn
?
TRIGGER:MAIN:VIDEO?
might return:
might return:
NTS;CH1;NEGA;NUMER;2;1;COLO;COLO;787;
59.94E+0;1050;2;890.0E-9;3.56E-6;15.00E-6;11.56E-6;
15.89E-6 as the current main video trigger parameters.
TRIGger:MAIn:VIDeo:BY
TDS 400A Option 05 Only
Sets or queries the video trigger delay mode. This is equivalent to using the
Video TV Delay Mode side menu.
Video TV Delay Mode side menu.
Trigger
TRIGger:MAIn:VIDeo:BY { TIMe | LINES | LINE }
TRIGger:MAIn:VIDeo:BY?
<Space>
?
TRIGger
:
:
TIMe
LINE
VIDeo
MAIn
BY
:
LINES
TIMe specifies a delay by time.
LINES specifies a delay by a number of video lines. For the TDS 400A, this
argument is available only for backward compatibility. If the TDS 400A receives
argument is available only for backward compatibility. If the TDS 400A receives
Group
Syntax
Examples
Group
Syntax
Arguments