Using the Menus
1740A/1750A Series Waveform/Vector Monitor User Manual
3-20
Table 3-1
.
Valid FORMAT selections.
525/60
625/50
BARS
SETUP
FORMAT
GBR
SMPTE / EBU
BETA
MII
GBR
SMPTE / EBU
BETA
MII
• •
75% 100%
• •
• •
• •
• •
• •
• •
• •
Y
N
[2]
[3]
75% 100%
Y
N
[1]
75% or 100% Bars selected through CONFIG—VECTOR menu.
Setup Yes or No is selected through the CONFIG—VECTOR
menu. The setup selection is only relevant with 525/60 BETA,
and is ignored in all other formats.
[2]
MII is identical to SMPTE/EBU in 625/50.
[1]
[3]
STANDARD
H
CONFIG—STANDARD
allows the user to set the cursor
unit of measure to
MV
,
IRE,
or
AUTO.
When
AUTO
is
selected, cursor units are automatically changed to
IRE when operating in the NTSC standard, and MV
when operating in the PAL standard.
H For dual-standard instruments, a menu selection of
NTSC
,
PAL,
or
AUTO
standard is also available.
When
AUTO
is selected, the operating standard is
changes automatically to match the standard of the
displayed input signal. A circled “A” precedes
NTSC or PAL on the readout.