Symptom 
Possible Cause 
Method of Correction 
Insufficient or varying power 
Check the power supply 
output and increase if 
necessary 
Insufficient pressure 
Ensure the pressure (an 
differential pressure) is high 
enough o operate the unit. 
Varying pressure 
Check the stability of the 
pressure regulation and 
improve if necessary. 
Ambient temperature too high 
or too low 
Place the unit in a suitable 
environment 
Output load resistance too low 
Ensure the voltmeter or 
data acquisition system or 
display has an impedance of 
greater than of 2.5kohm 
Gas temperature too high or 
too low 
Ensure the gas temperature 
is within the recommended 
operating range 
Inaccurate control 
Unit damaged or faulty 
Contact Omega Engineering 
Gas flow through unit not 
completely stopped 
Ensure there is no flow 
through the unit. The 
easiest way to do this is to 
plug both the inlet and 
outlet. 
Severe fluctuations in the 
ambient temperature e.g. unit 
in direct sunlight 
Carry out the rezero 
procedure in a stable 
environment 
Unstable power supply 
Check the stability and 
suitability of the power 
source 
Problems with 
rezeroing 
Insufficient warm-up period 
Allow the unit to warm-up 
for at least 5 minutes.