National Instruments 3.21E+04 Manual De Usuario

Descargar
Página de 181
Chapter 4     Signal Connections
 National Instruments Corporation
4-19
AT-MIO/AI E Series User Manual
Figure 4-5  summarizes the recommended input configuration for both 
types of signal  sources.      
Figure 4-5.  Summary of Analog Input Connections
Input 
Differential 
(DIFF)
Single-Ended —
Ground 
Referenced
(RSE)
Single-Ended —
Nonreferenced
(NRSE)
Floating Signal Source 
(Not Connected to Building Ground)
Grounded Signal Source
Examples
•  Ungrounded Thermocouples 
•  Signal conditioning with isolated outputs
•  Battery devices
Examples
•  Plug-in instruments with
   nonisolated outputs
+
-
+
-
V
1
ACH(+)
+
-
+
-
V
1
ACH(+)
ACH (-)
See text for information on bias resistors. 
See text for information on bias resistors. 
R
Signal Source Type
ACH (-)
AIGND
+
-
+
-
V
1
ACH
AIGND
+
-
+
-
NOT RECOMMENDED
ACH
+     V
    -
V
1
Ground-loop losses, V
g
, are added to
measured signal
+
-
+
-
V
1
ACH
AISENSE
R
AIGND
+
-
+
-
V
1
ACH
AISENSE
AIGND
AIGND