Chapter 4     Signal Connections
AT-MIO/AI E Series User Manual
4-40
 National Instruments Corporation
Figures 4-23 and 4-24  show the input and output timing requirements 
for the CONVERT* signal.
Figure 4-23.  CONVERT* Input Signal Timing
Figure 4-24.  CONVERT* Output Signal Timing
The ADC switches to hold mode within 60 ns of the selected edge.  This 
hold-mode delay time is a function of temperature and does not vary 
from one conversion to the next.  Separate the CONVERT* pulses by at 
least one conversion period.
The sample interval counter on the AT E Series board normally 
generates the CONVERT* signal unless you select some external 
source.  The counter is started by the STARTSCAN signal and 
continues to count down and reload itself until the scan is finished.  It 
then reloads itself in readiness for the next STARTSCAN pulse.
A/D conversions generated by either an internal or external 
CONVERT* signal are inhibited unless they occur within a data 
acquisition sequence.  Scans occurring within a data acquisition 
sequence may be gated by either the hardware (AIGATE) signal or 
software command register gate.
Rising-edge
polarity
Falling-edge
polarity
t
w
t
w
= 10 ns minimum
t
w
t
w
= 50-100 ns