NI PXI/PCI-5114 Specifications
Specification
Value
Comments
Spectral Characteristics
Spurious-Free 
Dynamic Range 
with Harmonics 
(SFDR), Typical
Range (V
pk-pk
)
10 MHz, 
–1 dBFS 
input signal
Includes the 
2
nd
 through 
the 5
th
 
harmonics
Measured 
from DC to 
125 MHz
20 MHz 
bandwidth 
limit filter off
All ranges except 0.04
0.04
58 dBc
58 dBc
Total Harmonic 
Distortion 
(THD), Typical
–58 dBc
–58 dBc
Effective 
Number of Bits 
(ENOB), 
Calculated*
7.2
6.2
Signal to Noise 
and Distortion 
(SINAD), 
Typical
44 dB
38 dB
RMS Noise 
Range
(V
pk-pk
)
20 MHz
Filter On 
20 MHz
Filter Off
50 
Ω 
terminator 
connected 
to input
All ranges 
except 0.04
0.28% FS
0.28% FS
0.04
0.28% FS
0.45% FS
*
ENOB = log
2
(sinad) – ½log
2
(1.5) – log
2
(A/V)
where
sinad = the linear representation of SINAD
A = amplitude of the supplied sine wave during the test
= (peak) full-scale range of the waveform recorder input
Refer to 1057-1994 IEEE Standard for Digitizing Waveform Recorders for information on equation derivation.