A&D AD-4212B-101/102/201/301 Manual De Usuario

Descargar
Página de 102
 
33
10-4  Description of the Class “Environment, Display” 
Condition ( 
Cond 
)  
Cond 0
 
This parameter is for sensitive response to the fluctuation of a weight value. 
Used when fast weighing speed is required. 
 
After setting, the balance displays FAST. 
 
Cond 2
 
This parameter is for stable weighing with slow response. Used to prevent a 
weight value drift due to vibration or drafts. 
After setting, the balance displays SLOW. 
Note 
For the AD-4212A series balance, with “Hold function (
Hold)” set to “ON 
(
1)”, this item is used to set the averaging time. 
 
Stability band width ( 5t-b )   
This item controls the width to regard a weight value as a stable value. When the fluctuation per 
second is less than the parameter, the balance displays the stabilization indicator and outputs or 
stores the data. The parameter influences the “Auto print mode”   
5t-b 0
 
This parameter is for sensitive response of the stabilization indicator. Used 
for exact weighing. 
 
5t-b 2
 
This parameter ignores slight fluctuations of a weight value. Used to prevent 
a weight value from drifting due to vibration or drafts. 
Note 
For the AD-4212A series balance, with “Hold function (Hold)” set to 
“ON (1)”, this item is used to set the stabilization range.
 
Hold function ( Hold ) (Animal weighing mode, AD-4212A only) 
This function is used to weigh a moving object such as an animal.   
When the weight value is over the weighing range from zero and the display fluctuation is within the 
stabilization range for a fixed period of averaging time, the processing indicator illuminates and the 
balance displays the average weight of the animal. When the animal is removed from the weighing 
pan, the display returns to zero automatically.   
This function is available only when the hold function parameter is set to “1” (the animal mode 
indicator ANIMAL illuminates) and any weighing unit other than the counting mode is selected. 
The averaging time and stabilization range are set in “Condition (Cond)” and “Stability band width (5t-b)”. 
Weighing range 
Averaging time 
Stabilization range
Cond 0
5t-b 0 
Cond 1
5t-b 1 
 
0.2 g or over 
Cond 2
2 seconds     Faster 
4 seconds        | 
8 seconds    More accurate  5t-b 2 
Small 
Big