36
Key mode B 
When the PRINT key is pressed or the external PRINT contact input is used, the balance outputs or 
stores the weighing data immediately regardless of the display condition. At this time, the display 
does not blink. 
 
Key mode C 
When the PRINT key is pressed or the external PRINT contact input is used, with the stabilization 
indictor on, the balance outputs or stores the weighing data immediately. If the stabilization indicator 
is not on, the balance waits for the indicator to turn on, and then outputs the weighing data. At this 
time, the display blinks one time. 
 
Interval memory mode 
The weighing data is periodically stored in memory.   
Required setting 
dout prt 
Interval memory mode 
 
dout data 
Data memory function is used. 
 
 
 
Stores weighing data. 
 
dout int Interval 
time 
Optional setting 
dout  
5-td1, 2, or 3  Adds the time and date. 
Example 
“For periodical weighing without a computer command and to output 
all of the data, to a computer, at one time” 
 
10-6  Description of the Item “Data format” 
 
A&D standard format  5if type 0   
This format is used when the peripheral equipment can receive the A&D format.   
If an AD-8121B is used, set the printer to MODE 1 or 2. 
  This format consists of fifteen or sixteen characters excluding the terminator. 
When numerical characters excluding a decimal point are exceeded eight characters for 
AD-4212B-102, the format becomes sixteen characters. 
  A header of two characters indicates the balance condition. 
  The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus 
sign is applied.   
  The unit, consisting of three characters, follows the data. 
  When comparison results are to be added (Cp fnc Cp-r 1), the results appear between the 
header and the data. 
 
 
 
 
 
 
 
 
 
 
CR LF
S T , + 0 0 0 . 1 2 7 8     g
S T
U S
O L
Q T
CR LF
S T , + 1 0 0 . 0 1 2 7 8
   
g
Header      Data (eight characters)      Unit 
Terminator 
Stable header 
Stable header of counting mode 
Unstable header 
Overload header 
Header Data 
Unit 
Terminator