SATO CX Series Printer
95
Calibrate Sensor
Calibrate Sensor
Calibrate Sensor
Calibrate Sensor
Command Structure
<ESC>CA
Example:
<ESC>CA
Placement:
This command should be sent after the sensor 
type has been selected with the <ESC>CI 
command. It should not be included with print 
job commands.
Default:
There is no default for this command.
Command Function
To adjust the label sensor for unique media and ribbon com-
binations.
Input to Printer
<ESC>A
<ESC>CI2
<ESC>CA
<ESC>Z
Printer Output
There is no printer output for this command. It will cause the 
printer to feed a label while the characteristics of the label 
backing or eye-mark are measured.
Special Notes
1. The sensor is calibrated for the currently selected sensor 
type (Eye-Mark or Transmissive).
2. If the Sensor is disabled with a <ESC>CI0 command, no 
calibration is performed.
3. The sensor must be re-calibrated whenever the sensor 
type is changed.
4. When calibrating the sensor for the CX Series TT Ther-
mal Transfer unit operating in the thermal transfer 
mode, the ribbon must be installed in the printer.