Pressure Systems, Inc.                                                                              98RK-1 & 9816 User’s Manual© 
 
 
 
Page 94 
 
www.PressureSystems.com 
ii 
dd 
Description 
'q' read 
index 
1B 00 
Set 
Thermal Update Scan Interval per eeee as decimal 
value (seconds), 1 <= eeee <= 3600 (1 hour) (default = 
60). 
11 
1C 
-- 
Reserved for factory use 
 
1D 00 
Set 
Minimum Source Air Pressure per eeee as decimal 
value (in psi) (default = 60).  
 
1E 00 
01 
Enable Source Air Check (default) 
Disable Source Air Check 
 
 
*** NOTE: Modification of option index ii = 13, 14, 16, 17, 18, 19, 1A, or 1B will result in that 
specified operating option immediately becoming the new module Power-On default.  However, 
for most other operating options to be retained after module power cycles, they must be 
explicitly saved when changed by function index ii = 07 to module Flash memory.  Likewise, the 
offset/gain coefficients calculated by calibration commands 'h', 'Z' or 'C', must be explicitly 
saved to transducer nonvolatile (EEPROM) memory  (see indexes ii = 08/09 in table above). 
 
The Valve Position indexes (ii=12 and ii=0C) each have two states (00/01) that when combined 
provide four (4) possible states of the C1/C2/C3/C4 internal valves.  This “Logical Rotary 
Switch” with four (4) positions is summarized in the following chart: 
 
    
 
 
C1 Energized 
C2 Not 
(0C=01) 
C2 Energized 
C1 Not 
(0C=00) 
 
C3 Energized 
C4 Not 
(12=01) 
 
 
PURGE 
position 
 
LEAK/CHECK 
position 
C4 Energized 
C3 Not 
(12=00) 
 
 
CAL/RE-ZERO 
position 
 
RUN 
position