SECTION 5.0
SOFTWARE CONFIGURATION
ITEM
CHOICES
FACTORY SETTINGS
CONFIGURE (continued)
C. Alarms (Section 5.7) (continued)
C. Alarms (Section 5.7) (continued)
3. Alarm 1 setup (TPC)
a. Setpoint
if oxygen (ppm)
-20 to 20 ppm
1 ppm
if oxygen (ppb)
-999 to 999 ppb
100 ppb
if oxygen (% saturation)
0 to 200%
100%
if chlorine, monochloramine, or ozone
-9999 to 9999 ppm
1.0 ppm
if pH
-2.0 to 20.00
7.00
if temperature
-5 to 130°C
25°C
b. Proportional
0 to 299.9%
100 %
c. Integral
0 to 2999 sec
0 sec
d. Derivative
0 to 299.9%
0.0%
e. Time period
10 to 2999 sec
30 sec
f. LRV (100% on)
if oxygen (ppm)
-20 to 20 ppm
0 ppm
if oxygen (ppb)
-999 to 999 ppb
0 ppb
if oxygen (% saturation)
0 to 200%
0%
if chlorine, monochloramine, or ozone
-9999 to 9999 ppm
0 ppm
if pH
-2.0 to 20.00
0.00
if temperature
-5 to 130°C
0°C
g. URV (100% off)
if oxygen (ppm)
-20 to 20 ppm
2 ppm
if oxygen (ppb)
-999 to 999 ppb
100 ppb
if oxygen (% saturation)
0 to 200%
200%
if chlorine, monochloramine, or ozone
-9999 to 9999 ppm
2 ppm
if pH
-2.0 to 20.00
2.00
if temperature
-5 to 130°C
100°C
h. Relay fault
None, open, or closed
None
4. Alarm 2 Control
a. Activation method
Oxygen, chlorine, monochloramine, ozone, temperature, pH
Oxygen
b. Control mode
Normal or TPC
Normal
5. Alarm 2 setup (normal)
a. Configuration
Low, high, or off
Low
Rest of alarm 2 setup is the same as alarm 1
6. Alarm 3 control and setup is the same as alarm 1
7. Alarm 4 control
7. Alarm 4 control
Alarm
Fault or off
Fault
8. Feed limit timer
a. Feed limit
Disable, alarm 1, alarm 2, or alarm 3
Disable
b. Timeout value
0 to 10,800 sec
600 sec
9. Interval timer
a. Select alarm
Disable, alarm 1, alarm 2, or alarm 3
Disable
b. Interval time
0 to 999.9 hr
24.0 hr
c. Repeats
1 to 60
1
d. On time
0 to 2999 sec
120 sec
e. Off time
0 to 2999 sec
1 sec
f. Recovery time
0 to 999 sec
600 sec
TABLE 5-1. Program Settings List (continued)
Continued on the following page
19