MODEL 54eA
SECTION 5.0
SOFTWARE CONFIGURATION
5.5 CHOOSING DISPLAY OPTIONS (CONTINUED)
8.  Set the remainder of the display parameters. Use the 
é and ê keys
to choose the desired parameter. Then press Edit (F4). Use the 
é
key to move the cursor to the desired selection. Press Save (F4) to
store. 
SECURITY CAUTION
The controller uses the timeout value to activate security. Once
the controller is unlocked by entering a security code, security will
not re-activate until a display timeout occurs. If timeout has been
turned off, security will never reactivate.
Temp units: °C
Output 1:  mA
Output 2:  mA
Exit
Edit
Language:  English
Display left: I
Display right:  Out 2
Exit
Edit
Display contrast:  40
Timeout:  On
Timeout value:  10 min
Exit
Edit
Temp units
°C or °F
Output 1
mA or % of full scale
Output 2
mA or % of full scale
Language
English, Français, Español, Deutsch, Italiano
Display left
sensor current (I), pH (appears only if free chlorine 
is being measured and pH was enabled), alarm 1 
setpoint (no units), alarm 3 setpoint (no units), 
or blank
Display right
sensor current (I), pH (appears only if free chlorine 
is being measured and pH was enabled), alarm 2 
setpoint (no units), alarm 3 setpoint (no units), 
output 2, or blank
Display Contrast  00 (lightest)-99 (darkest); the display contrast 
changes as the number changes
Timeout
Timeout returns the display from any other screen 
to the main display if no key is pressed before the 
timeout value is exceeded.
Polling address
Identifies controller in multi-drop HART applications.
Timeout value
29