Emerson Process Management 53eA Manual De Usuario

Descargar
Página de 116
74
4.02
pH
Buf1done: 
4.00 
@25°C
Abort
Cont
4.02
pH
Autocal buffer 1
Abort
Cont
Adjust temperature
Calibrate pH
Output trim
Exit
Enter
Buffer calibration
Standardize pH
pH slope
Exit
Enter
1.  Be sure the pH feature has been enabled. See Section 5.8.
2.  Obtain two buffer solutions. Ideally the buffer pH values should bracket the range of pH values to be meas-
ured.
3.  Remove the sensor from the process liquid. If the process and buffer temperatures are appreciably different,
place the sensor in a container of tap water at the buffer temperature. Do not start the calibration until the sen-
sor has reached the buffer temperature. Thirty minutes is usually adequate. 
13.2  AUTOMATIC TWO-POINT CALIBRATION
MODEL 54eA
SECTION 13.0
CALIBRATION - pH
4.  From the main display, press any key to obtain the main menu.
With the cursor on "Calibrate" press Enter (F4).
NOTE
If Hold was enabled in Section 5.6, the hold screen will appear.
To activate hold, refer to Section 5.6, step 11.
5. Press 
the 
ê key three times to move the cursor to "Calibrate pH".
Press Enter (F4).
6.  Press Enter (F4).
7.  Rinse the sensor with water and place it in buffer 1. Be sure the
glass bulb and reference junction are completely submerged. Swirl
the sensor. Press Cont (F3). "Wait" flashes until the reading stabi-
lizes. 
If the controller appears locked, the reading is not stable enough.
Investigate and correct the cause of the noise or drift (see Section
14.8). Alternatively, change the stability limits (see Section 5.8) and
repeat the calibration.  
8.
The screen at left appears once the reading is stable. Use the 
é or
ê key to change the reading to the nominal pH of the buffer. The
nominal pH is the pH of the buffer at 25°C. Press Cont (F3).