Trilogy200 clinical manual
112
Alarm
Priority
Audible
Visual (Alarm 
Indicator Button 
and Display)
Device 
Action
User/Clinician Action
High 
Respiratory 
Rate
High
• • •   • •
Red flashing button; 
“High Respiratory 
Rate” message
Operates
User: Press Alarm Indicator/
Audio Pause button to 
silence alarm. Or, press Left 
(Reset) button to reset alarm. 
Continue using device. If 
alarm continues, contact your 
health care professional.
Clinician: Verify patient 
status.
Low Respiratory 
Rate
High
• • •   • •
Red flashing button; 
“Low Respiratory 
Rate” message
Operates
User: Press Alarm Indicator/
Audio Pause button to 
silence  alarm. Or, press Left 
(Reset) button to reset alarm. 
Continue using device. If 
alarm continues, contact your 
health care professional.
Clinician: Verify patient 
status.
High  
Inspiratory 
Pressure
Escalates 
from 
Audible 
indicator 
to Medium 
priority and 
then High 
priority
 (for first two 
consecutive 
occurrences)  
• • • 
(for 3rd 
consecutive 
occurrence)  
• • •   • •
 
(for 10th 
consecutive 
occurrence)
Peak Pressure symbol 
turns red. When 
condition first occurs, 
a beep will sound. 
When condition 
occurs for third time, 
button flashes yellow 
and yellow “High 
Inspiratory Pressure” 
message appears. 
When condition 
occurs for 10th 
time, button flashes 
red and red “High 
Inspiratory Pressure” 
message appears.
Operates
User: Press Alarm Indicator/
Audio Pause button to silence 
alarm. Or, press Left (Reset) 
button to reset alarm. Make 
sure tubing is not pinched or 
blocked. If alarm continues, 
contact your health care 
professional.
Clinician: Verify patient 
status. If problem continues, 
have device serviced.