Philips Trilogy 200 Manual De Usuario
Trilogy200 clinical manual
112
Alarm
Priority
Audible
Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
High
Respiratory
Rate
Respiratory
Rate
High
• • • • •
Red flashing button;
“High Respiratory
Rate” message
“High Respiratory
Rate” message
Operates
User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status.
status.
Low Respiratory
Rate
Rate
High
• • • • •
Red flashing button;
“Low Respiratory
Rate” message
“Low Respiratory
Rate” message
Operates
User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status.
status.
High
Inspiratory
Pressure
Inspiratory
Pressure
Escalates
from
Audible
indicator
to Medium
priority and
then High
priority
•
(for first two
consecutive
occurrences)
• • •
(for 3rd
consecutive
occurrence)
• • • • •
(for 10th
consecutive
occurrence)
Peak Pressure symbol
turns red. When
condition first occurs,
a beep will sound.
When condition
occurs for third time,
button flashes yellow
and yellow “High
Inspiratory Pressure”
message appears.
When condition
occurs for 10th
time, button flashes
red and red “High
Inspiratory Pressure”
message appears.
Operates
User: Press Alarm Indicator/
Audio Pause button to silence
alarm. Or, press Left (Reset)
button to reset alarm. Make
sure tubing is not pinched or
blocked. If alarm continues,
contact your health care
professional.
Audio Pause button to silence
alarm. Or, press Left (Reset)
button to reset alarm. Make
sure tubing is not pinched or
blocked. If alarm continues,
contact your health care
professional.
Clinician: Verify patient
status. If problem continues,
have device serviced.
status. If problem continues,
have device serviced.