Trilogy200 clinical manual
126
4.  Connect the patient circuit. Fit the ventilator-end of the flexible 
tubing onto the bacteria filter.
Preventive Maintenance
Your new Philips Respironics Trilogy ventilator includes a blank PM 
Service Due label affixed to the bottom of your device.  The label is 
blank, which allows you to document when service is due.  Service 
is due every 10,000 hours or 24 months, whichever comes first 
depending on the usage of the device.  
Service Date
Service Hours
 
You will need to record both the date and hours on the blank label 
to maximize your in-service interval.  Record the service date as 24 
months from the date first placed on a patient.  Record the hours as 
10,000 hours for initial usage.  
Use the device blower hours to determine when service is required.   
The device blower hours are listed in the device software in the 
information menu.
Installing a New Bacteria 
Filter
Preventive Maintenance 
Sticker