Trilogy200 clinical manual
144
D.  Verify Reset
 
Wait 40 seconds and verify the following auto-reset conditions:
 
The High Priority audible indicator has stopped 
sounding
 
The red light on the Alarm Indicator/Audio Pause 
button has stopped flashing
E.  Restore Ventilator Settings
 
Modify the ventilator settings and change the following value 
shown in Table 8.
Setting
Value
Low Vte/Low Vti
Off
4.  Verify Circuit Disconnect Alarm
This procedure verifies that the Circuit Disconnect alarm is working 
properly.  It assumes that you have attached the test lung, verified 
ventilator settings, and turned on ventilator power as described in 
the Initial Setup section.
A.  Change Circuit Disconnect Ventilator Setting
 
Modify the Circuit Disconnect ventilator setting to match the 
value shown below in Table 9.
Setting
Value
Circuit Disconnect
10 seconds
B.  Disconnect Test Lung
 
Disconnect the test lung from the circuit.
C.  Verify the Alarm
 
Wait approximately 10 seconds and verify the following alarm 
signals:
 
The High Priority Audible Indicator sounds
 
A red light flashes on the Alarm Indicator/Audio 
Pause button
Table 8 - Restore Ventilator 
Settings
Table 9 - Ventilator Settings
Note: 
 The Low Inspiratory or 
Low Expiratory Pressure Alarm 
may also be detected.