Trilogy200 clinical manual
154
 
Measured Patient Parameters
Parameter
Range
Accuracy
Vte/Vti
0 to 2000 ml
Greater of 15ml or 15% of reading
Minute Ventilation
0 to 99 l/min
Calculation based on measured Vte or Vti 
and Respiratory Rate
Estimated Leak Rate
0 to 200 l/min
N/A
Respiratory Rate
0 to 80 BPM
Greater of 1 BPM or 10% of reading
Peak Inspiratory Flow
0 to 200 l/min
3 l/min plus 15% of reading
Peak Inspiratory Pressure
0 to 99 cm H
2
O
Greater than 2 cm H
2
O or 10% of reading
Mean Airway Pressure
0 to 99 cm H
2
O
Greater than 2 cm H
2
O or 10% of reading
% Patient Triggered Breaths
0 to 100%
N/A
I:E Ratio
9.9- 1: 1-9.9
Calculation based on Inspiratory Time and 
Expiratory Time
All flows and volumes are measured at BTPS conditions.
Spontaneous Breathing During Power Failure Conditions
Flow Set 
Point (l/min)
Inspiratory Resistance (cm H
2
O)
Expiratory Resistance (cm H
2
O)
Active 
Circuit
Active Circuit with
Proximal Flow 
Sensor
Passive 
Circuit
Active 
Circuit
Active Circuit with
Proximal Flow 
Sensor
Passive 
Circuit
30
< 3.0
< 3.5
< 2.0
< 2.0
< 2.0
< 1.5
60
< 9.0
< 10.5
< 5.0
< 3.0
< 4.5
< 4.0
Disposal
Dispose of this device and batteries in accordance with local regulations.