FOR CLINICIAN’S USE ONLY 
 
Accessing Prescription Setting Screens
Full Menu Access Mode
The ventilator has two levels of menu access, Full and Limited. Full 
Menu Access allows you to alter all available settings. Limited Menu 
access permits the user to alter only those prescription parameters 
that affect patient comfort, such as Rise Time, Flex, and Ramp Start 
Pressure, if those parameters are available as part of the prescription. 
The ventilator defaults to Full Menu Access mode. 
After accessing the Main Menu, if the device is in Limited Menu 
Access mode, you can use the following Setup key sequence to enter 
Full Menu Access mode and be able to change prescription settings:
 
Press the Down button and the Alarm Indicator/Audio 
Pause button simultaneously for several seconds. This will 
temporarily place the device in Full Menu Access mode.
When you perform this key sequence from the Monitor screen, 
the Main Menu screen appears and an audible indicator sounds 
indicating you are now in Full Menu Access mode.  
When you perform this key sequence when the airflow is off, the 
Setup screen appears and an audible indicator sounds. 
You can go into the Options menu and permanently change the 
Menu Access setting to Full Menu Access. Otherwise, the device 
will return to the Menu Access mode stored in the setting once you 
exit the menu screens or if one minute passes without pressing any 
device buttons.  If you are in the Setup mode, and an SD card is in the 
device, “Write Event Log to SD Card” will appear in the menu.
Note: The Full Menu Access key 
sequence can be performed 
either from the Power Off screen 
or from the Monitor screen.
Note:  When the airflow is off 
and AC power is removed from 
the device for more than 5 
minutes, the device will enter 
a low power mode to save 
battery life.  When the device is 
in low power mode, the Setup 
key sequence will be ignored.  
Press the Start/Stop button, or 
connect AC power, or insert an 
SD card to exit the low power 
mode.
Note:  Philips Respironics 
recommends that after you are 
finished changing prescription 
settings, and you give the 
device to the patient, you set 
the device back to Limited 
Menu Access mode so 
patients cannot change their 
prescription settings.
 WARNING
The information on this page is ONLY for health care professionals. Remove this page from the manual 
before giving the manual to the patient!