Trilogy200 clinical manual
10
Patient Circuit
Exhalation valves, patient circuits, and water traps are shipped clean, 
not sterile. Cleaning and disinfection of these parts should follow 
individual institution processes and conform to guidelines provided by 
Philips Respironics with each accessory.
External DC Power
Do not use the same external battery to operate both the ventilator 
and any other equipment such as power chairs.
An external battery should only be connected to the ventilator using 
the Philips Respironics Trilogy External Battery Cable. This cable is 
fused, pre-wired, and properly terminated to ensure safe connection 
to a standard deep-cycle, lead acid battery. Use of any other adapter or 
cable may cause improper operation of the ventilator.
The ventilator should only be connected to an automotive electrical 
system using the Philips Respironics Trilogy Automotive Adapter 
(when available). This adapter is fused, filtered, and designed for safe 
connection to a standard automotive electrical system. Use of any 
other adapter or cable may cause improper operation of the ventilator.
Do not operate the ventilator from a car electrical system when 
starting the vehicle or jump-starting the vehicle. Electrical transients 
during starting may cause improper operation of the ventilator.
Electrostatic 
Discharge (ESD)
Do not use antistatic or conductive hoses or conductive patient tubing 
with the device.
Notes
 
This product does not contain natural latex rubber or dry 
natural rubber in patient or operator accessible areas or 
in the air path or breathing circuit.