Chapter 3 Modes, Features, and Alarms
21
Therapy Mode Table
The following table summarizes all of the therapy modes and the settings available in each mode. Some settings in the table 
are dependent upon other settings. For instance, if the circuit type is set to Active with Flow, then the Flow Trigger Sensitivity, 
Leak Compensation, and Flow Cycle settings will display.
Note: Pressure Support, referred to in the table below and later in this manual, is defined as  IPAP - EPAP or Pressure - 
PEEP (PC-SIMV).
Therapy Modes
CPAP
S
S/T
T
PC
PC-SIMV
CV
AC
SIMV
Dual Prescription
Ther
ap
y P
ar
amet
ers
Circuit Type
CPAP
IPAP
AVAPS (On, Off)*
IPAP Max Pressure
IPAP Min Pressure
EPAP
Pressure
Pressure Support (PS)
PEEP
Tidal Volume
Breath Rate
Inspiratory Time
Trigger Type*
Flow Trigger Sensitivity
Leak Compensation
Flow Cycle Sensitivity
Ramp Length
Ramp Start Pressure
Flex *
Rise Time
Flow Pattern
Sigh
Circuit Disconnect
Apnea
Apnea Rate
High Vte*
Low Vte*
High Vti*
Low Vti*
High Minute Ventilation
Low Minute Ventilation
High Respiratory Rate
Low Respiratory Rate
High Inspiratory Pressure
Low Inspiratory Pressure
* Trigger Type, AVAPS, and Flex settings are only available with the Passive circuit type. High Vte and Low Vte settings are only 
available for Passive and Active Flow circuit types. High Vti and Low Vti settings are only available with the Active PAP circuit 
type. Leak compensation is only available with the Active Flow circuit type.