Philips Trilogy 200 Manual De Usuario

Descargar
Página de 181
Chapter 4 Ventilator Setup
53
Detachable Battery Pack
Philips Respironics offers a detachable Lithium-Ion battery pack. To 
use the detachable battery pack, snap the battery into place on the 
back of the ventilator, as shown in the following illustration. When 
the device is not connected to an AC power source or an external 
battery, the detachable battery will power the device, if attached.  
The length of time the ventilator will operate on battery power 
depends on many factors such as device settings, battery charge 
level, and condition or age of the battery. When fully charged, a new 
battery can power the ventilator for approximately three hours under 
typical patient conditions.
Whenever the ventilator is connected to AC power, it will 
automatically recharge the detachable battery pack. A completely 
discharged detachable battery will reach 80% charge status within 8 
hours, when charging at approximately 23˚ C ambient temperature.
Insert and securely latch the detachable battery into the device as 
shown below.
Properly Installed
Detachable
Battery
Insert
Detachable
Battery
 CAUTION
The detachable and internal 
batteries wear out based on 
the amount of use (hours or full 
charge-discharge cycles). The 
battery capacity and life are also 
reduced by operation at higher 
temperatures.
 CAUTION
Only use the Philips Respironics 
Trilogy Detachable Battery with 
the device.
Attaching the Detachable 
Battery
 CAUTION
Prolonged operation or storage 
at elevated temperatures may 
reduce the service life of the 
detachable or internal battery 
and other internal components 
of the ventilator.